ABS main body;
● LCD display screen, easily setting, clear and intuitive;
● Air temperature display and servo control;
● Skin temperature display and servo control,
double skin temperature sensor;
● Humidity display and servo control;
● Display time and heater power;
● 24 hours Parametric trend diagram display;
● Audible & Visual Alarms: power failure, sensor failure,
higher & lower temp, air & skin temperature override,
fan failure, low water level, system failure;
● >37℃ Preset Mode: power failure reset memory,
alarm silence, self-test function;
● Hood with single slope, big front door, 6 access ports, 4 tube grommet;
● 2 Storage cabinets, Protective guard rail, IV poles;
● RS-232 Output Connector (use for network system);
● Optional Pulse Oximeter monitor system;
● Optional oxygen concentration measure and control system;
● Optional ECG system;
● Optional baby weight scale system;
● Audio, video, and network interfaces are reserved for data
and function extensions.
Technical specification:
1 . Air temperature
Air temperature display range: 0℃-60℃;
Air temperature control range: 20.0℃-37.0℃
37.1℃-39.0℃
Air temperature accuracy: ≤±0.3℃;
Air temperature deviation alarm: ±3..0℃
2 . Skin temperature
Skin temperature display range: 0℃-60℃;
Skin temperature control range: 20.0℃-37.0℃
37.1℃-38.0;℃
Skin temperature accuracy :≤±0.3
Skin temperature deviation alarm: ±1.0
3. Humidity
Humidity display range: 0%--99%RH;
Humidity control range: 30%--95%;
Humidity control accuracy:
(10% - 90% at 20 - 40℃)±5%RH.
4. Oxygen concentration (Optional)
Oxygen concentration setting range: 21%-65%
Oxygen concentration display range: 18%-99%
Oxygen concentration display accuracy: 2%
Oxygen concentration control accuracy: 3%
5. Baby weight scale(Optional)
Weight range: 500g-10kg
Weight accuracy: ±10g
Display trend diagram of baby weight
6. Pulse Oximeter monitor (Optional)
SpO2: measure range 0-100%, display resolution 1%;
Pulse rate (PR): measure range 25-250bpm,
display resolution 1bpm;
Perfusion index(PI): measure range 0-20%,
display resolution 0.001%;
Variability index: measure range 0-100%,
display resolution 1%.
7. ECG (Optional)
ECG measure range: 0.15mV-5.5mV
ECG measure accuracy: 2.36uV/LSB
Breath measure range: 15-120rpm
Heart rate measure range: 15-300bpm
Heart rate measure accuracy: ±1bpm
Body temperature measure range: 0-50℃
Body temperature measure accuracy: 0.1℃